ZEISS 3 Product and Functional Description | 3.4 Software DescriptionUser Privilege DescriptionChange Toolbar Enables the user to change the toolbar.Change User Direc-toryEnables the user to change the location where all user specific param-eters and configurations are saved.Extractor Enables the user to change the extractor voltage.Gun Align Enables the user to modify the alignment of the electron beam.Gun Off Enables the user to switch off the field emission filament.Stage Initialise Enables the user to initialize the motorized stage.Supervisor Enables the user to perform the following actions:§ Start the Administrator, create and edit users§ Set User Max EHT§ Modify the filament current§ Set up, edit, and delete global stage coordinates§ Save common macros and toolbars§ Save common recipes§ Activate Partial Vent on Standby, Z Move on vent, Protect Z,Go to HV@Shutdown, EHT Off & Log Off, and Leave GunON at Shutdown.§ Use the bakeout functionVent Enables the user to vent the specimen chamber.3.4.1.4 SmartSEM Program SuiteThe SmartSEM Program Suite comprises the EM server, which implements the internal communi-cation between control software and microscope hardware, plus several programs and utilities.The main purpose of the SmartSEM Program Suite is to access all necessary microscopy parame-ters and software features to capture SEM data and optimize image acquisition.Access: Windows start menu > SmartSEMProgram DescriptionChamberScope Enables you to display the chamberscope image and the detector im-age at the same time.OptiProbe Calibra-tionEnables you to calibrate the OptiProbe option.ReadMe Contains important information on the currently installed version.Release Notes Contains an overview of all SmartSEM versions including new devel-opments and specific details.RemCon32 Serial interface for remote operation via RS232, e.g. for EDXLicense: REMCONSampleHolder-GalleryEnables you to inspect the dimensions of all possible specimen holdersas well as to set the dimensions of the custom specimen holders.Enables you to activate the available specimen holders for SmartSEM.Instruction Manual ZEISS GeminiSEM series | en-US | Rev. 2 | 349500-8138-000 65