Part NumberIssueDatePage61599388801301/2018113 / 249113 (249) 01/20188.3.2 CVI3 calibration functionTo perform the calibration, the system must be connected as follows:To connect the Delta and the CVI3, use the same cable to connect the Delta to the external barcodereader, and a crossed serial cable to be connected to the serial port of the CVI3.NOTE: For CVI3 connection, CVI3 calibration function details and CVI Monitor software,please refer to the specific documentation.To execute the calibration, perform the following steps: Connect the system as shown in the figure above (the PC and the CVI3 must belong to the samenetwork). Ensure that the SIMAP-Box interface is disabled.NOTE: Refer to the paragraph “Enabling the SIMAP-Box” for further details about how toenable/disable the SIMAP-Box. On the Delta, define a Tool and a Pset associated to it, with the torque (and possibly angle) totest the CVI3 tool. The batch size defines the number of tests used for the calibration. Do notenable the barcode in the Pset. Start the CVI Monitor software on the PC connected to the CVI3.CVI3Delta 6D/7DReferencetransducer (i.e.DST, DRT) withmechanical jointsimulatorDelta serialcableCrossedserial cableNetworkPC withCVI Monitor