EW/EG-N/EWB-e-1627 23Turn the adjusting knob on the right side of thebalance to the WEIGH position.Calibration is now finalised.The balance will automatically return to weigh-ing mode.5.8.3 Adjusting with external weight (only KERN EW and EWB)Carry out adjustment with the recommended adjusting weight (see Chapter 1 „Tech-nical Data“). The adjustment can also be carried out with different adjusting weights(see table), but not ideal from a metrological point of view. Information concerningthe adjusting weights is available at: http://www.kern-sohn.comModel Alternative adjusting weightEW 220-3NM 100 gEW 420-3NM 100 gEW 620-3NM 200 gEW 820-2NM 200 gEW 2200-2NM 500 gEW 4200-2NM 1000 gEW 6200-2NM 2000 gEW 12000-1NM 5000 gEWB 220-2M 100 gEWB 620-2M 200 gEWB 1200-1M 500 gCALbuSYEnd.