EW/EG-N/EWB-e-16275610 TroubleshootingThe balance should be switched off for a short time following an interruption in theprogramme sequence and disconnected from the mains supply. It is then necessaryto repeat the weighing process from the beginning.Help:Interruption Possible causeWeight display is not illuminated. • The balance is not switched on.• The mains supply connection has beeninterrupted (mains cable not pluggedin/faulty).• Power supply interrupted.The weight display changes continu-ally• Draught/air movement• Table/floor vibrations• The weighing plate is in contact with for-eign matter.• Electromagnetic fields / static charging(choose different location/switch off inter-fering device if possible)The weighing result is obviouslyincorrect• The balance display is not set to zero• Adjustment is no longer correct.• Great fluctuations in temperature.• Electromagnetic fields / static charging(choose different location/switch off inter-fering device if possible)Switch the balance off if other error messages should appear and then switch onagain. Contact the manufacturer if the error message does not disappear.