EW/EG-N/EWB-e-1627265.10 Verification switch and official sealPrior to verification the verification switch must be moved from the displayed position(see direction of arrow) to verification position. Once in this position, a parenthesiswill be displayed around the last display point.Following calibration the balance is sealed at the marked position.Balance calibration is not valid without the “official seals“.Position of the “official seals“:Move verification switchBalances that are subject to compulsory verification must be taken out of op-eration if:• The weighing result of the balance is outside the error limit. Therefore, apply aknown test weight (approx. 1/3 of the nominal load) to the balance at regular in-tervals and compare with the display value.• The reverification deadline has been exceeded.Official sealOfficial seal