EW/EG-N/EWB-e-1627386.2.3 Percent weighing(excl. model KERN EW120-4NM)Display symbol: %Percent weighing allows the weight to be displayed as a percentage in relation to areference weight. The displayed weight value is adopted as a fixed prescribed per-cent value. (Standard setting: 100%).Operation Display1. Use the key to switch the bal-ance on. Use the key to select unitchange-over [ % ] (see chap. 6.2.1).Information:Tare containers can also be used duringpercent weighing. Before starting percentweighing use the key to tare out thecontainer.2. Operate the key.[P. SEt] will flash on the display.3. Place the reference weight = 100% inthe balance dish.4. Operate the key.The reference weight is stored.ο 0 I I I F0 %P. SEt %P. SEt %ο 0 I I I F100.00 %