EW/EG-N/EWB-e-1627346.2.2 Piece counting(excl. model KERN EW120-4NM)Display symbol: PCSWhen counting units you can either count items into or out of the container. In orderto be able to count a substantial number of items, the average weight per item mustbe determined using a small quantity (reference units).The greater the reference unit, the greater the counting accuracy.A particularly high reference must be chosen for small or greatly varying parts.The work cycle is divided into four stages:• Tare the weighing container• Determine the reference unit• Weigh in the reference unit• Count the itemsOperation Display1. Use the key to switch the balanceon. Use the key to select unitchange-over Pcs(see chap. 6.2.2).2. Tare containers can also be usedduring unit counting. Before starting unitcounting use the key to tare out thecontainer.3. Operate the key.The reference unit will flash on the dis-play.ο 0 I I I F Pcs0ο 0 I I I F Pcs0PcsM on 10