EW/EG-N/EWB-e-1627528.3 Interface descriptionThe output format, output control, transfer speed and parity bit can be set followingthe choice of a certain operating mode. The various possibilities are described inchap. 7.2.2 “Parameters for the serial interface “.8.4 Data Output8.4.1 Data Transmission FormatsBy selecting the corresponding function at your balance one of the two following dataformats can be set:• 6-digit data format(excl. model KERN EW-120-4NM)Consisting of 14 words including the final signal; CR=0DH, LF=0AH (CR=returntravel / LF=line feed)1 2 3 4 5 6 7 8 9 10 11 12 13 14P1 D1 D2 D3 D4 D5 D6 D7 U1 U2 S1 S2 CR LF• 7-digit data format1 2 3 4 5 6 7 8 9 10 11 12 13 14 15P1 D1 D2 D3 D4 D5 D6 D7 D8 U1 U2 S1 S2 CR LFNote: The 7-digit format is identical with the 6-digit format except for the addition-al signal D8.8.4.2 Algebraic signP 1 = 1 wordP 1 Code Meaning+ 2 B H Data are 0 or positive- 2 D H Data are negativesp 20 H Date are 0 or positive8.4.3 DataD 1 to D 7 7 words with 6-digit format (excl. model KERN EW-120-4NM)D 1 to D 8 8 words with 7-digit formatD * Code Meaning0 - 9 30 H – 39 H Data 0 to 9 (max. 6 characters in 6-type format). (Point) 2 EH Decimal point, position not fixedSp 20 H Space character, leading zero oppressed