EW/EG-N/EWB-e-1627 477.2 List of the function parametersThe balance has been set to a certain standard configuration in the factory. This ismarked by a ☆.Function Display Choice Description of the choicepossibilitiesBar graph 1 b.G. 0 Off☆1 OnTolerance weighing(excl. EW-120-4NM)2 SEL ☆0 Off1 On (chap. 7.2.1)Zero alignment 3 A.0 0 No zero point correction☆1 Automatic zero point correction activated.Automatic shutoff after 3 min.for rechargeable battery op-eration (function is only avail-able inrechargeable batterymode)4 A.P. 0 Automatic shutoff deactivated for re-chargeable battery operation (optional).☆1 Automatic shutoff activated for rechargea-ble battery operation (optional).Display speed 5 rE. 0 Setting for metering1 Sensitive and fast2 ↓☆345 Non-sensitive but slowVibratory filter 6 S.d. 1 Sensitive and fast (very tranquil installationlocation).☆2 ↓34 Non-sensitive but slow (very unsettledinstallation location).5 only EW-120-4NM6 only EW-120-4NMInterface 7 I.F. 0 Interface not active(excl. EWB models) ☆1 6-digit data format (excl. EW-120-4NM)2 7-digit data format (chap. 7.2.2)Weight unit(only selectable, if the cali-bration switch is not in thecalibration position seeCh. 5.10)81↓85S.u.1☆01 (g)2☆14 (ct)15 (oz)16 (lb)17 (ozt)18 (dwt)19 (grain), (excl. EWB models)1A (tl Hong Kong)1b (tl Singapore,Malaysia)1C (tl Taiwan)1d (mom)1E (to)3☆20 (Pcs) excl. EW-120-4NM4☆IF (%) excl. EW-120-4NM5☆00 Unit not set (can not be set at 81.S.u.)