160 Keysight N5990A MIPI M-PHY Receiver Test User Guide5 HS TestsParametersThe parameters used in Expert Mode for this test, are as follows:• CDR Loop Bandwidth (if CDR is selected in the Special Parameterdialog, see “BERT Analyzer Settings”)• CDR Peaking (if CDR is selected in the Special Parameter dialog, see“BERT Analyzer Settings”)• CDR Transition Density (if CDR is selected in the Special Parameterdialog, see “BERT Analyzer Settings”)• CDR Loop Order (if CDR is selected in the Special Parameter dialog,see “BERT Analyzer Settings”)• Target BER• IBerReader Init Mode• Test Sequence• Re-Init sequence after Reset DUT• Lane Under Test Termination Model• Tested Accumulated Differential Voltage• Tested Accumulated Voltage Range• Set OffsetThese parameters are listed in alphabetical order in Table 6.DependenciesThe calibrations required for this test are as follows:• Reference Clock Calibration• Level Calibration Terminated• Accumulated Voltage calibration TerminatedResultsAn example of a HTML Viewer for the Test 2.1.2-HS-RX AccumulatedDifferential Input Voltage Tolerance Test procedure is shown in Figure 97.The results comprise of the following:• The common parameter list (in Expert Mode)• A test data table (Refer to Table 21)