Integrator's Guide26© 2017 Nortek AS5.6 SETAVG/GETAVG/GETAVGLIMSet/get averaging mode settings and get the relevant limits.Argument DescriptionNC Number of cellsCS Cell Size (m)BD Blanking Distance (m)CY Coordinate System (“ENU”, ”XYZ“, “BEAM”)PL Power Level [dB](-100 dB to switch off transmit-20.0 dB to 0.0 dB)AI Average interval (s)VP UnusedVR Velocity range along beam [m/s]DF Data Format0 – AD2CP format 1.1 –Legacy Aquapro format2 - AD2CP format 23 - AD2CP format 3NPING Number of pingsNB Number of beams(Select number of beams, 0 select beams according the PLAN,VD setting)CH Beam selection(Select beams, 0 select beams according the PLAN,VD setting. Example: 134select the three beams 1, 3 and 4)MUX Multiplexor Selection0 - 1,0, ping all beams in parallel1, ping beams in sequenceBW Bandwidth selection. (“NARROW”, “BROAD”)ALTI Enable altimeter in AVG measurementsBT Enable bottom track measurement in AVG samplingICE Enabled ice velocity measurement in AVG samplingALTISTART Altimeter start of measurement distance [m]ALTIEND Altimeter end of measurement distance [m]RAWALTI Storage of Raw Altimeter dataThe actual valid range for the various parameters for the firmware version is used can be found byusing the GETAVGLIM command. This command has the same arguments as the SETAVG/GETAVGcommands shown in the list above. The output format for limits is described in Data Limit Formats