Multimeter Mode - The "Trigger", "Horizontal" and "Display" MenusV - 110 SCOPIXMultimeter Mode (cont’d)The "Trig" MenuSource/Level Selection of trigger type and level on each channel. Triggering takes place if acondition described by a line of the “Trigger” table is verified.The trigger level should be defined in the channel measurement dynamic.Trigger leads to the logging of the instant and trigger characteristics.The events logged can be accessed from the "Disp." Æ "Defaults" menu.Source The channel number.Type The type of trigger for each channel.There are several possible types:¾ None (no triggering)¾ Lower than¾ Higher than¾ Lower/Higher¾ ExteriorIn "Multimeter" mode, several conditions are monitored simultaneously.Level 1 Adjustment of the main trigger level using the stylus.Level 2 Adjustment of the auxiliary trigger level using the stylus.This tab is enabled only if the "External" trigger type is selected.min duration > The fault will be evidenced if the fault condition defined by the type andlevels is present for a parameterisable period from 480 ms to 670 ksaccording to the selected recording duration (48 ms to 670 ks if yourinstrument is equipped wth the EXTENDED ACQUISITION MEMORYoption).www.ShopAEMC.comShop for AEMC products online at: 1.877.766.5412