XS series operating manualXS user manual v0.55 - 13 -Permanent current measurement Direct reading of the current on a shunt resistance in the test circuit. Display of the value on a 999 points digital indicator. Accuracy : (2.5% + 2 U) of the read value (1U=0.01mA).In VDC accuracy is given for resistance loads > 1 M.Fault indication By message on the LCD display, red and green LED, and audible signal (possibility to inhibit). Memory storage of the breakdown voltage. Memory storage of the leakage current in IMAX mode. Cut off of the high voltage at the first zero crossing of the control sinusoidal signal of the HVtransformer, thus without over voltage.Timer(mode) MANUALThe timer is not used during the test . The voltage output is manually controlled by pressing on theUP/DOWN arrows. The test stops when a breakdown occurs or when pressing on the red pushbutton on the front panel.(mode) DEFAULTThe timer is not used during the test . The voltage output is defined in the parameter VOLTAGE line.The test stops when a breakdown occurs or when pressing on the red push button on the frontpanel.(mode) AUTOThe test is defined by three consecutive times during which the output voltage rises linearly up to thepreset value ( RISE time) , then is hold to the preset value ( HOLD time) and finally bring back tozero ( FALL time).The RISE and FALL times can be set from 0 to 1 sec. by 0.1 sec. step and from 1 to 999 sec. by1 sec. step.The HOLD time can be set from 1 to 999sec. by 1 sec. step.