XS series operating manualXS user manual v0.55 - 65 -Activation of the capacitive sample detectionWARNINGThis parameter line only concerns the SXS16.Fig 8.11It allows the activation of the capacitive sample detection before a dielectric strength test starts. Bydefault, the capacity threshold is set to 10nF.Parameter memorizationXS series unit allows to store in 10 memories (numbers 0 to 9) measurement parameters (voltage,threshold, time , ....). From the measurement menu, to modify the memory number:Press on the [MEM:x] function keyDisplay shows :Fig 8.12With the UP/DOWN arrow keys, increase or decrease the memory number (from 0 to 9). Theparameters recall line (area 3 of the LCD screen) indicates the content of each memory.Enter the selected memory either by pressing again on the [MEM:x] function key or by pressing theENTER key.From the parameter menu, to modify a memory number:Press on the [MEM:x] function keyWith the UP/DOWN arrow keys, increase or decrease the memory number (from 0 to 9). The inputparameters lines indicate the content of each memory.Enter the selected memory either by pressing again on the [MEM:x] function key or by pressing theENTER key.WARNING