K0430 Issue No. 2 16Leak Test (LEAK TEST) TaskA - External pressure source B - Pressure regulator C - Device/system under test• If required, use the INPUT key to change pressure units.• Set-up the leak test WAIT and DURATION times to the required values as shownbelow. A minimum wait period of 3 minutes is recommended.• Close the vent valve and pressurize the device/system to the required LEAK TESTpressure.• Press the RUN (F2) key to start the leak test. When completed, the beeper soundsand the leak test results are written to the display.ABPRESSURE INT barF2F1RUNCHANGEVALUEWAITDURATIONSTART PRESSSTOP PRESSPRESS CHANGELEAK RATEsecssecsbarbarbarbar/m6060max 30VTASK : LEAK TESTSNAPSHOT MODEDPI 615 ISCTaking MeasurementsTest Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176TestEquipmentDepot.com