19 K0430 Issue No. 2GeneralAdvanced task allows the user to configure the instrument to monitor one of a number ofdifferent input measurements and outputs (sources). Additionally, five process functions,Tare, Max/Min, Filter, Flow and % Span can be applied to the input functions.Select InputTo select an input channel, select ADVANCED task from the task menu. The displayshows the list of the input selections and, if available, the PROCESS soft box (F1) and theUNITS soft box (F2).The following procedure shows the method of input channel selection and the method ofchanging units:Note: Left/right arrow keys function as page up/down keys.Refer to pages 20 to 23 for details of process functions.Ambient Temperature MeasurementTo set-up the instrument to read ambient temperature, proceed as follows:Note: Ensure that the temperature reading has stabilised.SELECT UNITSSELECT AIRTEMPERATUREINPUTSELECT INPUTSELECT INPUTPROCESS UNITSUNITSCELSIUSFAHRENHEITAMBIENT TEMPNO INPUTAdvanced Task Select Input