K0430 Issue No. 2 56(6) Apply the full-scale pressure to the external transducer and store the full-scale(FS) point.(7) Release the applied pressure and disconnect the pressure reference. Press theENTER key to accept the calibration. Press the EXIT key once to return to thecalibration menu or four times to quit the CALIBRATION and SETUP modes.(8) Check the calibration of the external transducer by applying the test pressures asdetailed in Table 1, page 42.Add External SensorTo add an external sensor to the directory of external sensors, proceed as follows:• Connect the required external transducer to the EXT TRANSDUCER socketlocated on the rear of the instrument.Digitally compensated transducers will be downloaded into the directory as soon as theF1 key is operated. Data for other types can be edited by selecting EDIT (F2). If the datacannot be edited a warning message (INVALID ACTION) is displayed.CALIBRATING PRESSURE EXTRANGE 20000 kPa gCALIBRATING PRESSURE EXTRANGE 20000 kPa gCALIBRATING PRESSURE EXTRANGE 20000 kPa gApply nominal FS valuePress ENTER key whenapplied value is stable +CHANGE VALUETest Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176TestEquipmentDepot.com