19Installation3.4.2 Remove the shipping transport locksPrior to system start-up, the shipping transport locks must beremoved from the sc analyzer.CAUTIONThe enclosure may tip forwards if it has not been fixed inplace. Only open the enclosure if the enclosure is properlymounted.Important note: The electrode insulation for the measuring unit isNOT a transport lock. Do NOT remove the cover on the measuringunit.1. Open the enclosure door and secure with the door hook.2. Remove the transport lock on the analyzer panel (Figure 9).Note: If the instrument is operated with a Filter Probe sc, it is equippedwith an internal compressor.3. Remove the cable tie and pull out the compressor transportlock to the left (Figure 10).Note: Keep the transport locks for transporting and storage.Figure 9 Remove the analyzer panel transport locks1 Transport lock