88Plumbing and Connection OptionsA.9 Option 5 plumbing and connectionsOption 5 uses an sc analyzer as a 2-channel analyzer with twoFILTRAX (FILTRAX 1 and FILTRAX 2), supplying two continuoussample streams. The waste from the analyzer and both FILTRAX isdischarged back into the drain through the optional heated drainhose LZY302 (230V) or LZY303 (115V).Refer to Figure 33 and the following instructions for Option 5:1. Install both FILTRAX into the sample stream. Refer to theFILTRAX User Manual for more information.2. Feed the heated hose from FILTRAX 1 through the analyzeropening (Figure 33, item 7). Use Seal Plug #1 to secure.3. Feed the heated hose from FILTRAX 2 through the analyzeropening (item 6). Use Seal Plug #1 to secure.4. Feed the heated drain hose through the analyzer (item 5). UseSeal Plug #1 to secure.Note: The two samples lines of the heated drain hose are not used.5. Connect the heated drain power connections. Refer tosection 3.5.5, page 27.6. Connect the heated drain tube to the T-fitting (item 8).7. Connect the sample line from FILTRAX 1 to the bottom inlet onOverflow Vessel 1 using the fittings (item 11).8. Connect the sample line from FILTRAX 2 to the bottom inlet onOverflow Vessel 2 using the fittings (item 12).