26Installation3.5.3 Insert tubing and/or cables3.5.4 Connect the Filter Probe to the analyzerCAUTIONThe enclosure may tip forwards if it has not been fixed inplace. Only open the enclosure if the enclosure is properlymounted.DANGERDisconnect power from the sc analyzer at the sc1000 beforeremoving the protective covers in the analyzer.Important Note: Make sure that for the use of a filter probe, thelevel of the water in which the filtration probe is submersed, isbelow the level of the analyzer.1. Open the enclosure door and secure with the door hook.2. Open the analyzer panel.3. Remove the two screws from the protective cover and removethe cover (item 1, Figure 14, page 24).4. Connect the earth ground (green/yellow) wire (item 9,Figure 16, page 28) from the Filter Probe sc to the groundterminal (item 5, Figure 16, page 28).5. Connect the power connector to the appropriate terminalconnection (items 4 and 11, Figure 16, page 28).6. Remove the three screws securing the bottom panel cover.(item 2, Figure 14, page 24). Remove the panel.7. Connect the data connector (item 10, Figure 16, page 28) tothe main board (item 12, Figure 16, page 28).8. Install all covers and panels.9. Connect the white air tubing (item 8, Figure 16, page 28) fromthe Filter Probe sc to the air tubing connection on the analyzer(Figure 16, page 28).10. For sample and drain line connections refer to A.4 , page 79.1 Guide the tubing or cables throughthe enclosure openings(Figure 15).2 Push the plug from the top onto thetubing or onto the cable.3 Pull down the plug with the tubingor the cables. Seal any unusedentry with Plug #3.