27Installation3.5.5 Connect the optional heated drainCAUTIONThe enclosure may tip forwards if it has not been fixed inplace. Only open the enclosure if the enclosure is properlymounted.DANGERDisconnect power from the sc analyzer at the sc1000 beforeremoving the protective covers in the analyzer.Refer to Figure 16 and the following procedure to connect theheated drain.1. Open the enclosure door and secure if necessary.2. Open the analyzer panel.3. Remove the protective cover (Figure 14, page 24).4. Connect the earth ground wire (green/yellow) to the groundwire terminal strip.5. Connect the cables for the heated drain (item 6, Figure 16,page 28) to the terminal block (item 3, Figure 16, page 28).6. Connect the drain tube as described in the appropriate OptionConfiguration, refer to section A.5 , page 80 for moreinformation.7. Place the drain tube to the appropriate drain or basin.8. Install all covers and panels.