84Plumbing and Connection OptionsA.7 Option 3 plumbing and connectionsOption 3 uses an sc analyzer with the FILTRAX. The waste fromthe analyzer is discharged back into the drain through the optionalheated drain hose LZY302 (230 V) or LZY303 (115 V).Refer to Figure 31 and the following instructions for Option 3:1. Install the FILTRAX into the sample stream. Refer to theFILTRAX User Manual for more information.2. Feed the heated hose from the FILTRAX through the analyzeropening (Figure 31, item 5). Use Seal Plug #1 to secure.3. Feed the heated drain hose through the analyzer opening(item 3). Use Seal Plug #1 to secure.Note: The two samples lines of the heated drain hose are not used.4. Seal the remaining opening with Seal Plug #3.5. Connect the heated drain power connections. Refer tosection 3.5.5, page 27.6. Connect the heated drain tube to the sample outlet T-fitting.