90Plumbing and Connection OptionsA.10 Option 6 plumbing and connectionsOption 6 uses two sc analyzers with two FILTRAX (FILTRAX 1 andFILTRAX 2). Samples from both FILTRAX are going into Analyzer 1using the 2-parameter configuration. The heated drain hoseconnects both sc analyzers. The waste from both analyzers isdischarged into a drain through the heated drain hose.Refer to Figure 34 and the following instructions for Option 6:1. Install both FILTRAX into the sample stream. Refer to theFILTRAX User Manual for more information.2. Install the first sc analyzer (Analyzer 1):a. Feed the heated hose from FILTRAX 1 through theanalyzer (Figure 34, item 25). Use Seal Plug #1 to secure.b. Feed the heated hose from FILTRAX 2 through theanalyzer (item 27). Use Seal Plug #1 to secure.c. Feed the heated drain hose through the analyzer (item 28).Use Seal Plug #1 to secure. Connect the heated drainpower connections. Refer to section 3.5.5, page 27.d. Remove the pre-installed drain tube attached to the valveblock and remove the T-fitting from the drain tube. Retainfor reuse with Analyzer 2.e. Connect the heated drain tube to the valve block connector.f. Connect the sample line from FILTRAX 1 to the bottom inleton Overflow Vessel 1 using the fittings (item 20).g. Connect the sample line from FILTRAX 2 to the bottom inleton Overflow Vessel 2 using the fittings (item 19).h. Change the analyzer to the 2-parameter configuration.Refer to Connect a 2-parameter option, page 78.i. Connect Sample Line 1 from the heated drain to OverflowVessel 1. Connect Sample Line 2 from the heated drain toOverflow Vessel 2.3. Install the second sc analyzer (Analyzer 2)a. Feed the heated drain hose from Analyzer 1 throughAnalyzer 2 (item 13). Use Seal Plug #1 to secure.b. Feed the heated drain hose through the analyzer (item 8).Use Seal Plug #1 to secure. Connect the heated drainpower connections. Refer to section 3.5.5, page 27.c. Seal the remaining opening with Seal Plug #3.d. Cut 25 mm from the drain tube that was removed fromAnalyzer 1. Connect the 25 mm piece of tubing to theT-fitting on Analyzer 2. Connect the other end of the tubingto the T-fitting that was removed from Analyzer 1. Refer toFigure 28, page 79 for T-fitting removal.e. Connect the drain tube from Analyzer 1 and the drain tubefrom Analyzer 2 to the T-fitting.4. Connect Sample Line 1 from Analyzer 1 to Overflow Vessel 1using the fittings (item 16). Connect Sample Line 2 fromAnalyzer 1 to Overflow Vessel 2 using the fittings (item 17).