DMTA-10022-01EN, Rev. C, January 2015Chapter 61206.7.7 Generating ReportsThe 45MG can generate inspection data reports without having to be connected to acomputer or printer. The following reports are available:File summaryShows basic statistics for the file (minimum thickness and location, maximumthickness and location, and high and low alarm conditions along with the mean,median, and standard deviation).Minimum/Maximum summaryShows a list of ID number locations that have the minimum and maximumthickness in a file.Alarm summaryShows a list of all the ID number locations where a low alarm and high alarmoccurred.File comparisonAllows you to select two files and compare them. The first file contains theprevious inspection data, and the second file contains the current inspection data.The report indicates the maximum wall loss and any area of greater wallthickness (growth), and their ID number locations.Minimum ReviewAllows you to select a file, and then review all the minimum thickness locations inthe file. In addition to verifying the thickness at all minimum locations, you canreplace these minimum thickness locations if necessary.To generate a report1. In the measurement screen, press [FILE], and then select REPORT.2. In the submenu, select the desired type of report. If you select:• FILE SUMMARY, go to step 3• MIN/MAX SUMMARY, go to step 4• FILE COMPARISON, go to step 5• ALARM SUMMARY, go to step 6• MIN REVIEW, go to step 73. In the FILE SUMMARY screen (see Figure 6-39 on page 121):a) Select the file for which you want to create the report.