DMTA-10022-01EN, Rev. C, January 2015Maintaining and Troubleshooting the 45MG 2057. When DUAL XDCR TEST is selected, in the DUAL XDCR TEST screen (seeFigure 12-3 on page 205):a) Press [MEAS] to start the test that measures the time of flight through eachside of a dual element transducer.The measured parameter values appear.b) Interpret the Tx and the Rx values as follows:• Similar values indicate a normal dual element transducer.• A difference between the values indicates that the delay line wear is notthe same for each element.• A missing value indicates that a cable is broken, or that the element is notfunctioning.c) Take note of the calculated ZERO VALUE.d) Press [ENTER] to terminate the dual element transducer test.Figure 12-3 The DUAL XDCR TEST screen12.6 Performing the Software Diagnostic TestThe software diagnostic (SW DIAG) function generates an error log documenting allerrors that occurred during instrument operation. Olympus uses this information totroubleshoot the 45MG operating software.To access the software diagnostic1. In the measurement screen, press [SETUP], and then select SP MENU.