DMTA-10022-01EN, Rev. C, January 2015Custom Setups for Single Element Transducers 169• IF BLANK in modes 2 and 3 only (see “Interface Blank” on page 180 fordetails)• ECHO 2 DETECT in modes 2 and 3 only (see “Detection of Echo 1 andEcho 2” on page 179 for details)• ECHO 1 DETECT (see “Detection of Echo 1 and Echo 2” on page 179 fordetails)• ECHO WINDOW (see “Detection of Echo 1 and Echo 2” on page 179 fordetails)• MB BLANK (see “Main Bang Blank” on page 175 for details)• TDG SLOPE (see “TDG Slope” on page 175 for details)• INIT GAIN (see “Initial Gain” on page 175 for details)• MAX GAIN (see “Maximum Gain” on page 174 for details)• PULSER POWER (see “Pulser Power” on page 172 for details)• PROBE TYPE• MEAS TYPE (see step 5.c in “Creating a Custom Setup for a Single ElementTransducer” on page 165 for details)4. Use the [] or [] keys to select the value for the selected parameter.5. Repeat steps 3 and 4 to adjust other parameters.6. Press [GAIN/WAVE ADJ] again to hide the waveform adjustment parameter.10.3 Detection ModesThere are three detection modes (Mode 1, Mode 2, and Mode 3):Mode 1Uses a direct contact transducer to measure the time of flight between the mainbang and the first back-wall echo (see Figure 10-3 on page 170).