LSM 510 QUICKSTARTLSM 510 META Switching on the System Carl ZeissB 45-0008 e 10/02 4-74.2 Switching on the SystemThe LSM system is turned on with the REMOTE CONTROL switch. This switches all the systemcomponents on except for the "Enterprise" UV laser.If the UV laser shall be used, it can be switched on after the start of the WINDOWS ® NT operatingsystem - but must always be switched on before the LSM 5 software is started.If REMOTE CONTROL switch is not used, turn the system on with the "I" button on the laser module;in addition, the jumper plug supplied must be connected to the POWER REMOTE CONTROL terminal.• Turn the REMOTE CONTROL switch to "ON"position (see Fig. 4-1).− This switches the entire system on.− Microscope and laser will be ready foroperation after a short time.− Computer boots up.− Computer hardware system test runs.Drive "A" of the computer must notcontain a floppy disk.The monitor shows a dialog box for selection ofthe operating system version.Fig. 4-2 Selecting the operating system versionFig. 4-1 REMOTE CONTROL switch PreviousNext |