Model 190 Installation, Technical and Operation8400-M022-O1 Rev J Page 634.11 Display High Resolution WeightIMPORTANT! The Hires (Hires?) mode requires a load of10% of Capacity on the scale.Hi resHigh Resolution Weight (from SETUP Prompt)1. If displaying High Resolution Weight is desired, with the SETUPprompt displayed, press UNITS/ key until the display shows theHires prompt.2. Place a calibrated test weight on the scale and press the TARE key.3. The display will change to show the amount of the test weight in"high resolution" mode (1/10 division), an h will be displayed to theright of the annunciators and the annunciators will alternately flashoff and on.4. Press the Fn/ key to increase the span by 0.1 division OR pressthe UNITS/ key to decrease the span by 0.1 division.5. Press the TARE key to exit the Display High Resolution Weightand advance to LoCoUt? prompt.Hires?High Resolution Weight (after TARE pressed at FSPAN prompt)1. With Hires? displayed, press the TARE key.2. The display will change to no.3. Place a calibrated test weight on the scale.4. Press the Fn/ key to toggle to yes and then press the TARE key.5. The display will change to show the amount of the test weight in "highresolution" mode (1/10 division), an h will be displayed to the right ofthe annunciators and the annunciators will alternately flash off and on.6. Press the Fn/ key to increase the span by 0.1 division OR press theUNITS/ key to decrease the span by 0.1 division.7. Press the TARE key to exit the Display High Resolution Weight andadvance to the LoCoUt? prompt.