MGR10 A/B/C - Revision : I (for 7.0 software revision) 121.7 Common specification to all model numbers1.7.1 Temperature compensationThe accuracy of the temperature measurement is ±0.1% and uses a standard Pt100 sensor. Thisaccuracy does not include errors due to the sensor itself. The temperature measurement range is0…+40C and the resistance measurement is referenced to 20C when this option is used.Accuracy of the MGR10-04 sensor : +/- ( 0.2% + 1°C)1.7.2 Lead resistanceA maximum lead resistance of 0.5 / Imes ( = 0.05 for 10A and 500 for 1mA ) ispermissible in each of the four measuring leads.1.7.3 Mains supplyUniversal power inlet Mains : 115/230V~ 10% single phase 47 to 63 Hz. Power consumption : 40VA without load, 70 VA maximum. Switching from 115 to 230 volts by switch in the line input plug. Temporized fuse protective in rear panel- 2AT for 230V 4AT for 110V. Battery powered with built in charger ( MGR10A version) The two batteries used are 5Ah, and the following performance can be expected fromfully charged batteries:o Digital :11 hours continuous use, with backlight on, and IEEE and PLC cardsfitted.o Analogue: Over 6000 simple measurements in FAST mode, using 10Ameasuring current. (Or approximately 3000 measurements using SLOWmode).The battery condition is constantly monitored, and if it is low, a warning symbol is displayed .Inthis situation operate as follows : Connect a mains cable on the unit rear mains socket Connect the mains cable on the mains Switch the rear panel switch in the ON position The FAST red Led lamp should light on to indicate the battery charging mode When the red Led lights off , the charge is over