MGR10 A/B/C - Revision : I (for 7.0 software revision) 404.8 Measure/Hold functionThis key triggers measurement cycles according to the trigger mode ( Single or Continuous – referto section 4.3)In Continuous mode : (this mode is not available when operating on batteries – MGR10A ) Press on [MES.] to start permanent measurement cycles , the « Meas » display is blinking ateach measurement Press again on [MES.] to store the last measured value on the display. Display indicates“HOLD” state .NOTA : this state doesn’t interrupt the measurement current , the current continue to flow throughthe specimen under test. Press again on [MES.] to resume the memorization of the last measured value and to displaynew measurementIn Single mode : Press on [MES.] to start a measurement cyclesThis stops the measurement current at the end of the cycle Press again on [MES.] to get the following measurementWARNING : when measuring with a 10A current , it is possible to create small sparks duringconnection and disconnection of the specimen . To avoid damaging the specimen under test , pleasemake the connection and disconnection without current flowing .6MES.MEAS.