MGR10 A/B/C - Revision : I (for 7.0 software revision) 6629.657 K 29.657E+3 (30K range)106.45 m 106.45E-3 (200m range)If a query (?) command results in a “command” or “execution” error, the error value +9.90E+37 is sentto the output buffer.Input Message TerminatorsIn RS-232 mode, all messages sent to the instrument must terminate with either a character or a character. It is also permitted to send both to terminate a message; the is ignored.For IEEE-488 mode, messages can be terminated in one of two ways;- by sending a character at the end of the message- by asserting the EOI line with the last character sentIt is also permitted to combine these, and assert EOI when sending the character.Message termination always forces the command path back to the root-level ready for the nextcommand message.Input BufferThe instrument receives messages into an input buffer and only starts executing commands after receiptof a message terminator. The buffer can store up to 100 characters including separators and terminator.Sending a new command before the existing command is executed may cause unreliable operation. It isadvisable to query the Status Byte (*STB?) to check on the current status of the microhmmeter beforesending a new command.Sending a command whilst the microhmmeter is transmitting (following a ? command) may result inthe transmitted response being corrupted. It is advisable to wait for the expected response to betransmitted before sending a new command.6.3.2 Configuration & Measurement CommandsConfiguration GroupThese commands allow the measurement configuration to be set up, but they do not trigger ameasurement.SENSe:FRESistance:RANGe <range>Sets the resistance measurement according to the following parameters. The current measurementis aborted.