MGR10 A/B/C - Revision : I (for 7.0 software revision) 14SECTION 2 : SETTING UP THE MGR102.1 Safety informationPlease read and follow these important safety instructions: Read the safety information at the beginning of this manual before operating the MGR10; Make the necessary electrical safety connection checks. In particular, select the correct line voltageand make sure that the correct fuse is installed. Incorrect voltage or fuse selection present both anelectrical safety and a fire hazard. Ensure that the rear panel fan is operating and that the vent is notblocked. When connecting to an electrical supply mains supply, the mains cord provided with the equipmentshould be used, and connected only to a mains supply with a suitable earth connection. Under nocircumstance should the equipment be operated with earth disconnected.2.2 Unpacking the instrumentWhen you unpack the MGR10, check that the following items are present before starting to use theunit: 1 x MGR10 Microhmmeter 1 x Mains Cord 1 x Operating manual 1 x Calibration CertificatePlease contact the Sefelec Customer Support Team immediately if any of these items are missing ordamaged.