60 CFS/CCS-BA-e-182514 Adjustment• Provide required adjustment weight, cf. chapter 1.The adjustment weight to be used depends on the capacity of the weight /counting system. Implement adjustment as close to the maximum load aspossible. Info about test weights can be found on the Internet at:http://www.kern-sohn.com• Observe stable environmental conditions. A warming up time (see chapter 1)is required for stabilization.• In order to avoid errors at the piece quantity determination, both balancesmust be adjusted with the same acceleration due to gravity.In case of non-compliance counting errors will result!14.1 Models CFS 300-3, CFS 3K-5Operation Display Switch-on balance and during the selftest press . „Pin“ Use the numeric keys to enter password:Either input four times zero „0000“ as a standardpassword, or input the user-defined password (input cf.parameter „Pin“ chapter 13). Confirm input by .„Pin“ „----„ Select bulk material or reference balance via .The appeared [] displays the current balance.When using as counting system, the bulk materialbalance as well as the reference balance must beadjusted. The adjustment process must be carried out onboth balances„tECH“ „LoCAL“„tECH“ „rEmotE“ If necessary, at balance zero display using selectthe weighing unit [g / kg or lb], which shall be used foradjustment. The appeared [] displays the currentweighing unit.Acknowledge with .„tECH“ „Unit“