CFS/CCS-BA-e-1825 61 Ensure that there are no objects on the weighing pan.Wait for stability display (indicator [] atopexpires), then press . Carefully place required adjustment weight into the centerof the weighing plate on display „LoAd“.Wait for stability display, then press . After the adjustment the balance will carry out a self-test.Remove adjusting weight during selftest, balance willreturn into weighing mode automatically.In case of an adjustment error or incorrect adjustingweight the display will show an error message (), repeat adjustment process.14.2 Models CFS 6K0.1, CFS 15K0.2, CFS 30K0.5Operation Display Switch-on balance and during the selftest press . „Pin“ Use the numeric keys to enter password: Either input four times zero „0000“ as a standardpassword, or input the user-defined password (input cf.parameter „Pin“ chapter 13). Confirm input by .„Pin“ „----„ When using as counting system, the bulk materialbalance as well as the reference balance must beadjusted. The adjustment process must be carried out onboth balances.Select bulk material or reference balance via .The appeared [] displays the current balance.Acknowledge with .„tECH“ „LoCAL“„tECH“ „rEmotE“ Using select the weighing unit [kg or lb] whichwill be used for linearization. The appeared [] displaysthe current weighing unit.Acknowledge with .„tECH“ „Unit“