68 CFS/CCS-BA-e-1825 Use the numeric keys to input password „9999“:Confirm input by . „Pin“ „----„ Select bulk material or reference balance via .The appeared [] displays the current balance.When using as counting system, the bulk materialbalance as well as the reference balance must beadjusted. The adjustment process must be carried outon both balances. Acknowledge with .„tECH“ „LoCAL“„tECH“ „rEmotE“ Using select the weighing unit [kg or lb] which willbe used for linearization. The appeared [] displays thecurrent weighing unit. Acknowledge with .„tECH“ „Unit“ Ensure that there are no objects on the weighing pan.Wait for stability display (indicator [] atopappears), then press . Carefully place first adjustment weight into the center ofthe weighing plate on display „LoAd 1“.Wait for stability display, then press . Carefully place second adjustment weight into thecenter of the weighing plate on display „LoAd 2“.Wait for stability display, then press . When „LoAd 3“ is displayed, place the third adjustmentweight carefully in the centre of the weighing plate.Wait for stability display, then press . After the adjustment the balance will carry out a self-test. The balance returns automatically into weighingmode. In case of an adjustment error or incorrectadjusting weight the display will show an error message( ), repeat adjustment process.