66 CFS/CCS-BA-e-182515.1 Models CFS 300-3, CFS 3K-5Operation Display Switch-on balance and during the selftest press . „Pin“ Use the numeric keys to enter password „9999“:Confirm input by . „Pin“ „----„ Select bulk material or reference balance via .The appeared [] displays the current balance.When using as counting system, the bulk materialbalance as well as the reference balance must belinearized. Both scales must be linearized.„tECH“ „LoCAL“„tECH“ „rEmotE“ If necessary, at balance zero display usingselect the weighing unit [kg or lb], which shall be used forlinearization. The appeared [] displays the currentweighing unit.Acknowledge with .„tECH“ „Unit“ Ensure that there are no objects on the weighing pan.Wait for stability display (indicator [] atopexpires), then press . Carefully place first adjustment weight into the center ofthe weighing plate on display „LoAd 1“.Wait for stability display, then press . Carefully place second adjustment weight into the centerof the weighing plate on display „LoAd 2“.Wait for stability display, then press . When „LoAd 3“ is displayed, place the third adjustmentweight carefully in the centre of the weighing plate.Wait for stability display, then press . Carefully place fourth adjustment weight into the center ofthe weighing plate on display „LoAd 4“.Wait for stability display, then press .