START-UPZEISS Mounting standard components Axiolab 564 430037-7444-001 05/20193.1.10 Mounting or replacing the 35 Whalogen lamp or the 10 W LEDilluminator for transmitted lightAxiolab 5 stands are equipped with a 10W LEDwhite light for transmitted illumination. The LEDilluminator can be alternatively changed to a 35WHalogen lamp.To insert or replace the halogen lamp/LEDilluminator, proceed as follows:Removing the cover• Switch off the microscope, remove the powercord on the microscope and allow it to cooldown at least 15 min.• Press the clamping jaw (Fig. 3-15/1) on thecover (Fig. 3-15/2) downwards. Swing the coverdown, remove from the retaining channels onthe stand and set aside.Changing the halogen lamp• Remove the lamp plug (Fig. 3-16/3) from thehalogen lamp (Fig. 3-16/2).• Press the loops (Fig. 3-16/1) on the securingclips of the lamp holder together and swingthem out to the front.• When changing the lamp, remove the old one(Fig. 3-16/2).• Position the new lamp with the lower frontedge between the contact surface and securingclips.• Lift the securing clips (Fig. 3-16/1) of the lamp holder with the lamp until they are completely enclosedin the lamp holder. In doing so, slightly press the ends of the securing clips together and guide pastthe two upper retaining elements. Release the pressure until the clamping bracket opens and engageson both sides in the retaining elements.• Check that the lamp is seated correctly and push the lamp plug (Fig. 3-16/3) onto the pins of the lamp(Fig. 3-16/2). Ensure that it engages properly to avoid bending the pins.• Insert the cable of the lamp plug into the stand so that it is not damaged when the cover is attached.Fig. 3-15 Removing the coverFig. 3-16 Changing the halogen lamp