CIRRUS HD-OCT User Manual 2660021169012 Rev. A 2017-12Electromagnetic Compatibility (EMC) 1- 11Guidance and manufacturer’s declaration – electromagnetic emissionsThe CIRRUS HD-OCT is intended for use in the electromagnetic environment specified below. The customer or user of the CIRRUS HD-OCTshould assure that it is used in such an environment.Emissions Test Compliance Electromagnetic environment – guidanceRF emissionsCISPR 11Group 1The CIRRUS HD-OCT uses RF energy only forits internal function. Therefore, its RFemissions are very low and are not likely tocause any interference in nearby electronicequipment.RF emissionsCISPR 11Class A The CIRRUS HD-OCT CIRRUS are suitable foruse in all establishments other than domesticestablishments and those connected to a lowvoltage power supply network which suppliesbuildings used for domestic purposes.Harmonic emissionsIEC 61000–3–2Class AVoltage fluctuations/flicker emissionsIEC 61000–3–3CompliesGuidance and manufacturer’s declaration – electromagnetic immunityThe CIRRUS HD-OCT is intended for use in the electromagnetic environment specified below. The customer or user of the CIRRUS HD-OCTshould assure that it is used in such an environment.Immunity Test IEC 60601 test level Compliance level Electromagnetic environment – guidanceElectrostatic Discharge(ESD) IEC 61000–4–2± 6 kV contact± 8 kV air± 6 kV contact± 8 kV airFloors should be wood, concrete, or ceramic tile. Iffloors are covered with synthetic material, therelative humidity should be at least 30%.Electrical fasttransient/burst IEC 61000–4–4± 2 kV for power supplylines± 1 kV for input/outputlines± 2 kV for power supplylines± 1 kV for input/outputlinesMains power quality should be that of a typicalcommercial or hospital environment.Surge IEC 61000–4–5 ± 1 kV differential mode± 2 kV common mode± 1 kV differential mode± 2 kV common modeMains power quality should be that of a typicalcommercial or hospital environment.Voltage dips, shortinterruptions, and voltagevariations on power supplyinput lines. IEC 61000–4–11<5% U T (>95% dip in U T )for 0,5 cycle40% U T (60% dip in UT ) for5 cycles70% UT (30% dip in UT ) for25 cycles<5% U T (95% dip in U T )for 5 sec<5% U T (>95% dip in UT )for 0,5 cycle40% UT (60% dip in UT ) for5 cycles70% U T (30% dip in U T ) for25 cycles<5% U T (95% dip in UT )for 5 secMains power quality should be that of a typicalcommercial or hospital environment. If the user ofthe CIRRUS HD-OCT requires continued operationduring power mains interruptions, it isrecommended that the CIRRUS HD-OCT be poweredfrom an uninterruptible source.Power frequency (50/60 Hz)magnetic fieldIEC 61000–4–83 A/m 3 A/m Power frequency magnetic fields should be at levelscharacteristic of a typical location in a typicalcommercial or hospital environment.UT is the a.c. mains voltage prior to application of the test level.