Checkout and Preventive Maintenance SL 800Service Manual Confidential and Proprietary SM-30-4072Order No. 2330-705 2 - 8 Version 02, 2019-10-24The printout may be obsolete. Check the SIS for the most current and approved version of this document.2.3.6 Create Log-FilesTools and aids: Empty USB flash drive with FAT32 formatProcedure:The SL 800 is able create a log file if required. It contains information on configuration, status ofindividual modules or any errors that may have occurred.1) Remove all USB cables from the power supply unit (any connections to a PC, fixation lamp andwide field illumination).2) Turn on SL 800 at the power supply unit.3) Insert the prepared USB drive into the lower USB-A port labelled with “USB 2A” at the powersupply. Writing of the SL 800 log file starts automatically. After a few seconds, the SL 800 confirmscompletion of log file creation by three beeps.4) Switch off the SL 800 at the power supply unit after the log life has been written successfully.5) Remove the USB flash drive.6) Reconnect the removed USB cables to the corresponding USB sockets at the power supply.7) Switch the SL 800 on at the power supply unit to continue working with the device.8) Check the log file for errors codes and plan further corrective maintenance measures, ifnecessary.