Troubleshooting Error Codes SL 800Service Manual Confidential and Proprietary SM-30-4072Order No. 2330-705 5 - 2 Version 02, 2019-10-24The printout may be obsolete. Check the SIS for the most current and approved version of this document.5.1 Log file structureThe SL 800 log file contains information on configuration, status of individual modules or any errorsthat may have occurred. The log file is very helpful for preparing a ZEISS service call in the event of atechnical problem with the device.General Information regarding serialnumbers and firmware versions.For all SL 800 components the serialnumber contains the materialnumber as well. From the 2 nd digit tothe 14 th digit represents the materialnumber, e.g. 0000002201092 forthe power supply.System status information for eachmodule, e.g. the currentmagnification position of theMicroscope.Error statistics and error codes foreach module.Use the specific error code and thelog file error code list for furthertroubleshooting, e.g. code 52 Check the cable connection at thePower Module,…Section for the current systemtemperatures.Fig. 44: Log file example