Troubleshooting Error Codes SL 800Service Manual Confidential and Proprietary SM-30-4072Order No. 2330-705 5 - 6 Version 02, 2019-10-24The printout may be obsolete. Check the SIS for the most current and approved version of this document.ErrorNo.Error Name Module Error description Error Correction827 E_ChrError_EEPROM_WRITE Microscope MicroscopeEEPROM errorsReplace the Microscope828 E_ChrError_EEPROM_ID Microscope MicroscopeEEPROM errorsReplace the Microscope834 E_ChrError_CHR_MOTOR_DRIVING_TIMEOUTMicroscope Cannot reach thedesiredmagnificationposition.Check all microscopecable connections,Replace the Microscope835 E_ChrError_CHR_MOTOR_FAULTMicroscope Defect microscopemotorCheck all microscopecable connections,Replace the Microscope836 E_ChrError_EEPROM_DEVICE_NOT_READYMicroscope MicroscopeEEPROM errorReplace the Microscope1219 E_PjrError_CAN1_TX_TIMEOUTIlluminationLED ModuleCAN-Bus error Replace the instrument1224 E_PjrError_EEPROM_INIT IlluminationLED ModuleIllumination LEDModule EEPROMerrorReplace the instrument1225 E_PjrError_EEPROM_TRIPLE IlluminationLED ModuleIllumination LEDModule EEPROMerrorReplace the instrument1226 E_PjrError_EEPROM_READ IlluminationLED ModuleIllumination LEDModule EEPROMerrorReplace the instrument1227 E_PjrError_EEPROM_WRITE IlluminationLED ModuleIllumination LEDModule EEPROMerrorReplace the instrument1228 E_PjrError_EEPROM_ID IlluminationLED ModuleIllumination LEDModule EEPROMerrorReplace the instrument1235 E_PjrError_LED_TMP_LIMIT_REDU2_EXCEEDIlluminationLED ModuleLED temperaturetoo high (>80 °C)Check the heat sink,replace the instrument1236 E_PjrError_LED_VLED_LIMITS IlluminationLED ModuleVLED-Limits not inspecificationCheck all cableconnections within theIllumination Module,replace the instrument1238 E_PjrError_LED_TMP_SENSOR_FAULTIlluminationLED ModuleLED temperaturesensor defect(temperature >100°C)Check all cableconnections within theIllumination Module1239 E_PjrError_EEPROM_DEVICE_NOT_READYIlluminationLED ModuleIllumination LEDModule EEPROMerrorReplace the instrument