113Expanded programs5. Press Verification.6. Insert cell adapter A (Figure 3, page 19) in cell compartment(2).7. Remove any remaining cells from the cell compartment.8. Close the cell compartment cover.9. Press Start.10. Follow the directions on the display and set various filtersone after another in the order displayed.11. Press Next after inserting a filter.After the last reading the results are displayed.12. Press PC & Printer icon to send the data to a USB memorystick, PC or to a printer.The files will be stored automatically as CSV files (CommaSeparated Value). The file name will be formatted as"Verification.csv".6.8.3.2 Test solutionThe test solution (refer to Section 9, page 137) serves as aregular quality control for the photometric accuracy of or therelevant spectral photometer.When results exceed allowable tolerances (given in the qualitycontrol certificate to the test record), contact the manufacturer.