134TroubleshootingError copying from USB memorystick. Error during update Start the procedure again or contact themanufacturer or his representative.Error in FTP connection. FTP error Make sure that the instrument isconnected to the network.FaultProgram not accessible.Please update program data!Barcode test not present Update program dataFaultClean cuvette!The cuvette is dirty or there areundissolved particles in the cuvetteClean the cuvette; allow the particles tosettleFaultTest program stopped!Please check lampClose the lid.Error [xx]Test program stops when the instrumentis startedCheck the lamp and replace it ifnecessary.Close lid.Press Start Again.FaultTest program stopped!Please remove the cuvetteClose the lid.Test program stops when the instrumentis startedRemove the cuvette/sample cell fromthe cell compartment.Press OK.ErrorSelfcheck stopped.Hardware error.Error [x]Electronic defect Contact technical support and indicatethe error numberErrorToo much ambient light!Move instrument into shadeor close the lidThe instrument sensors detects toomuch ambient light.Decrease ambient light. (Avoid directsunlight.)Close lid.No help function is available for thisprogram.Shelf life exceeded!Use chemicals?The analysis is possibly erroneous. Usenew chemicalsNo evaluation! Error in the test database / userdatabaseCheck programmingContact technical supportNo barcode! No barcode foundReinsert the cell.If the barcode is not recognised, contactthe manufacturer or his representative.No instrument backup present! Check the USB memory stick.No valid data for these parameters! Data analysis not possible, nomeasurement data Change the selection.No valid data found! View Data not possible in data log Change the selection.No measurement data present! Data analysis settings cannot beconfigured without measurement data. Change the selection.Control range not reached! Data analysis limits not reached This is a warning notice. The controllimit set was not reached.Control range exceeded! Data analysis limits exceeded. This is a warning notice. The controllimit was exceeded.Concentration too high! Calculated concentration is higher than999999 Dilute sample and measure againOver measuring range The measured absorbance is above thecalibration range of the test Dilute sample and measure againError displayed Definition Resolution