AEJ-C/AES-C/PLJ-C-BA-e-1621 15929.6 Detailed information to communication logs29.6.1 ZeroingCommand: Z CR LFPossible replies:Z˽A CR LFZ˽D CR LFCommand accepted, command executingCommand execution completedZ˽A CR LFZ˽^ CR LFCommand accepted, command executingCommand valid but zero reset range was exceededZ˽A CR LFZ˽E CR LFCommand accepted, command executingTime limit for stabilisation of weighing balance display wasexceededZ˽ CR LF Command valid but cannot be executed at the moment29.6.2 TaringCommand: T CR LFPossible replies:T˽A CR LFT˽D CR LFCommand accepted, command executingCommand execution completedT˽A CR LFT˽^ CR LFCommand accepted, command executingCommand valid but tare range was exceededT˽A CR LFT˽E CR LFCommand accepted, command executingTime limit for stabilisation of weighing balance display wasexceededT˽I CR LF Command valid but cannot be executed at the moment29.6.3 Query tare valueCommand: OT CR LFResponse:1 2 3 4 - 12 13 14 - 16 17 18 19O T ˽ Tare ˽ Unit ˽ CR LFTare value: - 9 characters right aligned