164 AEJ-C/AES-C/PLJ-C-BA-e-162129.6.11 Start continuous output in current weighing unitCommand: CU1˽CR LFPossible replies:CU1˽I CR LF Command valid but cannot be executed at the momentCU1˽A CR LF Command accepted, command executingWeighing values are shown in current weighing unit1 2 3 4 5 6 7-15 16 17-19 20 21S U I Stabilitysign ˽ Signs Weight ˽ Unit CR LF29.6.12 Stop continuous output in current weighing unitCommand: CU0˽CR LFPossible replies:CU0˽I CR LF Command valid but cannot be executed at the momentCU0˽A CR LF Command accepted, command executing29.6.13 Enter lower tolerance limitCommand: DH˽XXXXX CR LFPossible replies:DH˽OK CR LF Command executedES CR LF Syntax error (incorrect weight format)29.6.14 Enter upper tolerance limitCommand: UH˽XXXXX CR LFPossible replies:UH˽OK CR LF Command executedES CR LF Syntax error (incorrect weight format)