168 AEJ-C/AES-C/PLJ-C-BA-e-162129.6.27 List of available operating modesCommand: OMI CR LFPossible replies:OMI˽ CR LFN° designation CR LFN° designation CR LFOK CR LFCommand accepted, available operating modes are shownOMI˽I CR LF Command valid but cannot be executed at the momentn Operating mode1 - Weighing2 - Parts counting3 - Percent determination4 - Dosing5 - Make up recipes6 - Animal weighing7 - Density determination “solid matter“8 - Density determination “liquids“9 - Maximum value function10 - Totalization11 - Tolerance weighing12 - Statistics13 - Pipette calibration29.6.28 Set operating modeCommand: OMS CR LFPossible replies:OMS˽OK CR LF Command validOMS˽E CR LF No parameter or incorrect formatOMS˽I CR LF Command valid but cannot be executed at the moment29.6.29 Query current operating modeCommand: OMG CR LFPossible replies:OMG˽n˽OK CR LF Command accepted, number of current operating mode will beshownOMG˽I CR LF Command valid but cannot be executed at the momentn Number of operating mode, see chap. 29.6.27