AEJ-C/AES-C/PLJ-C-BA-e-16219717.2.2Enter known reference single weight as numeric valueSelect application , see chap.16.1. The reference singleweight last used will be shown in the informational line of text. At the first start-up as value 0.0000g.Press the function key assigned to the function (Display example F2).Enter known reference single weight with the help of the navigation keys andthe active digit will be flashing.Confirm input by pressing thePRINTbutton.The saved reference single weight will be shown in the informational line oftext. Tare if necessary, place weighing good and read off the number of items.•The reference single weight will remain saved until you define a newreference.•Return to weighing mode, see chap.17.2.1