AEJ-C/AES-C/PLJ-C-BA-e-1621 161Example:S CR LF Sent command: Send stable weighing value instandard weighing unitS ˽ A CR LF Command accepted, command executingS˽˽˽˽ -˽˽˽˽˽˽ 8 . 5 ˽g˽˽CR LF Command executed,Weighing value is shown in default weighing unitStability symbol*? instable (3Fh, 0x3F)˽ stable (20h, 0x20)29.6.6 Send weighing value immediately in standard weighing unitCommand: SI CR LFPossible replies:SI˽I CR LF Command valid but cannot be executed at the moment1 2 3 4 5 6 7-15 16 17-19 20 21S I ˽ Stabilitysign ˽ Signs Weight ˽ Unit CR LFExample:S I CR LF Sent command, see chap. 29.4S I˽?˽˽˽˽˽˽˽1 8 . 5˽k g˽CR LF Command executed,Weighing value is shown in default weighing unit