AEJ-C/AES-C/PLJ-C-BA-e-1621399.3Settings Navigation in menu see chap.8.2Menu structure:Menu itemParametersAvailable settings / explanationP2.2.1READINGP2.2.1.1FilterEXTRA FASTWeighing balance reacts very sensitively and very fast,very stable place of installation.FASTWeighing balance reacts sensitively and fast,stable place of installation.MEDIUMStandard – normal ambient conditionsSLOWWeighing balance reacts insensitively but slowly, unstableplace of installation.VERY SLOWWeighing balance reacts very insensitively, but very slowly,very unstable place of installation.P2.2.1.2Weighing speedPRECISERest position control precise, unstable place of installationFASTRest position control fast, very stable place of installationFAST + PRECISERest position fast + precise, stable place ofinstallationP2.2.1.3AutozeroSelectable yes / noP2.2.1.4Last digitALWAYSLast decimal place will be shownNEVERLast decimal place will be hiddenWHEN STABLELast decimal place will only be shown for stableweighing valuesP2.2.1.5EnvironmentSTABLEUNSTABLE