AEJ-C/AES-C/PLJ-C-BA-e-1621359.1.3Weighing with tareTaringThe dead weight of any weighing container may be tared away by pressing a button,so that the following weighing procedures show the net weight of the goods to beweighed.•Place the weighing container on the weighing pan and close the draft shields,as required.•Wait until the stability display appears (), then press. Zero displayand indicator (Net) appear. The weight of the container is now internallysaved.•Weigh the material.•Wait until the stability display appears ().•Read net weight.•When the balance is unloaded the saved taring value is displayed withnegative sign.•Taring negative values is not permitted. Any attempt to tare negativevalues will result in error message “Err 3“•The taring process can be repeated any number of times. The limit isreached when the whole weighing range is exhausted.•You can assign the tare weight to a product in the database. Selectingthe product will automatically load the tare weight, see chap.15.2.1.Automatic tare functionWhen the automatic tare function is enabled the weighton the pan will first be saved as tare, see chap.9.3.6